The Agilent/Varian BA2 senTorr Vacuum Gauge Controller is a half-rack instrument engineered for continuous pressure measurement across rough to high vacuum ranges. This controller delivers precise monitoring and control of multi-gauge vacuum systems with individual LED digital displays per gauge tube, front-panel keypad operation, and configurable set points. Factory-selectable tube configurations accommodate diverse vacuum applications, from standard systems to ultra-high vacuum environments.
– Technical Specifications
Measurement & Display
• Pressure range: Rough to high vacuum
• Display: One LED digital display per gauge tube
• Set points: Four assignable set points (one assignable to any of three gauge tubes)
Gauge Tube Configurations
Factory-configurable to BA2, BA2C, CC2, CC2C, BA, CC, UHV, or UHV2c. The BA2 configuration includes two thermocouple gauges and one Bayard-Alpert gauge tube.
Electrical
• Input voltage: 90–127 VAC or 208–250 VAC, selectable, 50/60 Hz
• Gauge supply: +24 V (dc) ±5% with <±1% ripple, 0–1 A per channel, 25 W per channel
• Fuse protection: 1.5 A per channel with self-resetting PTC element
• High voltage: Up to 3000 V present in equipment
Data Retention
Parameter values retained for four years accumulated off-time during power loss.
Environmental
Meets specifications at 25°C (±5°C) and 90% relative humidity, non-condensing.
Physical
• Dimensions: 3.5" H × 8.0" W × 15" D (half-rack mount)
• Includes 6-foot power cord and fuse set
• Optional rack-mounting kits available for 19-inch standard racks
– Key Features
• User-friendly front panel with keypad lockout capability
• Optional RS232 or RS485/422 serial communication boards
• Analog outputs and remote control options on rear panel
• Resistive 12 Ω degas for Bayard-Alpert models; E-beam degas for UHV models
• Vacuum system grounding support rated for minimum 10 A
– Typical Applications
Process monitoring and control in vacuum deposition, outgassing analysis, leak detection, and ultra-high vacuum research environments requiring multi-point pressure surveillance.
– Compatibility & Integration
Integrates with Agilent/Varian gauge tube families and supports legacy BA and CC configurations alongside modern UHV platforms. Optional communication boards enable integration into SCADA and networked laboratory control systems.




















Reviews
There are no reviews yet.