The Agilent/Keysight 423A is a precision RF detector that converts RF power levels into proportional DC voltage values for accurate power measurements in 50 Ω coaxial systems. Operating bias-free across 10 MHz to 12.4 GHz, this point-contact detector delivers relative power measurements up to 100 mW with exceptional sensitivity and flat frequency response. The detector outputs negative polarity DC signals (positive available via Option 003) through a BNC connector, making it ideal for signal detection, swept-source leveling, and broadband component characterization.
– Technical Specifications
• Frequency Range: 10 MHz to 12.4 GHz
• Input/Output Impedance: 50 Ω nominal input; <15 kΩ output shunted by 10 pF
• Maximum Input Power: 100 mW peak or average; 100 mW typical for <1 minute short-term exposure
• Frequency Response: ±0.2 dB over any octave (10 MHz to 8 GHz); ±0.5 dB (10 MHz to 12.4 GHz)
• Sensitivity (25°C): High-level mode 0.4 mV/µW CW
• SWR: 1.20 (10 MHz–4.5 GHz); 1.35 (4.5–7.0 GHz); 1.50 (7.0–12.4 GHz)
• Temperature Coefficient: 0.015 dB/°C (0°C to 55°C)
• Operating Temperature: 0°C to +55°C
• Dimensions/Weight: 3 × 1 × 1 in; 1 lbs
– Key Features
• No DC bias required for operation
• Point-contact detector element with stable, repeatable response
• Dual sensitivity modes accommodate both high-level and low-level detection
• BNC output connector compatible with standard oscilloscope inputs
• Low-pass filtering recommended below 1 GHz to attenuate RF leakage through output
– Typical Applications
• CW and pulsed power detection
• Automatic level control and sweeper leveling
• Frequency response characterization of microwave networks
• Pulse envelope demodulation for oscilloscope observation
• Harmonic-frequency comparison as a mixer element
– Compatibility & Integration
The 423A accepts Type N or APC-7 inputs depending on frequency range. Option 001 supplies matched detector pairs with 11523A load resistor for optimized square-law performance. Option 002 provides matched load resistor alone. Standard negative output polarity integrates directly with conventional test setups; Option 003 enables positive-polarity operation when required.