The California Instruments OMNI 3-16 is a programmable AC power source engineered for laboratory and industrial simulation of AC power conditions across single and three-phase configurations. The unit delivers 16 A output and integrates PC-based control software with high-speed data acquisition for compliance testing to international standards including IEC 61000-3-2, IEC 61000-3-12, IEC 61000-3-3, and IEC 61000-3-11. This platform supports harmonic analysis up to order 200 (10/12 kHz) and flicker characterization, making it suitable for power quality assessment, equipment immunity qualification, and harmonics distortion testing.
## Technical Specifications
**Power Output**
• 16 A rated output
• Single and three-phase configurations available
• Output voltage: 0–156 Vrms L-N (low range) and 0–312 Vrms L-N (high range)
• Output frequency: DC, 40 Hz to 5 kHz; fundamental line frequency 45–65 Hz
**Measurement Capabilities**
• Harmonic analysis: up to harmonic order 200 (10/12 kHz); spectrum display to harmonic 40, expandable to 9 kHz
• Voltage measurement accuracy: 0.1% + 10 mV (harmonics); 0.1% + 0.1% per 100 Hz + 5 mV
• Current measurement accuracy: 0.1% + 5 mA (Phase-A); 0.15% + 5 mA (Phase-B & C); harmonic current 0.1% + 0.02%/100 Hz + 5 mA
• Power factor range: −1.000 to +1.000 (±0.003 accuracy)
• Power measurement accuracy: 0.15% + 0.1 W per phase
• Phase measurement accuracy: 50–2500 Hz, 0.2° + 0.2° per 100 Hz
**Control & Connectivity**
• PC-based Windows software for control, analysis, and reporting
• USB Type B interface
• SCPI programming support for remote operation
• High-speed data acquisition with ASCII format output to disk
## Key Features
• Harmonic analysis to order 200 and flicker measurement
• Multi-phase capability for three-phase system testing
• Precision voltage and current measurements across wide frequency span
• Extended frequency range options
## Typical Applications
• Power quality and harmonics compliance testing (IEC 61000-3-2, IEC 61000-3-12)
• Flicker assessment (IEC 61000-3-3, IEC 61000-3-11)
• Voltage dip and interruption immunity qualification per EN/IEC 61000-4-11
• Harmonics immunity testing per EN/IEC 61000-4-13
## Compatibility & Integration
The OMNI 3-16 integrates into larger test systems and operates within multi-unit configurations. Remote programming via SCPI enables automated test sequences. Data streams to PC storage for post-acquisition analysis.