Yokogawa AQ6360 1200-1650nm Optical Spectrum Analyzer,This optical spectrum analyzer delivers high-resolution measurements over the 1200 to 1650 nm wavelength range, ideal for telecommunications and photonics applications. It features a fast sweep speed, narrow resolution bandwidth, and low polarization dependency. The graphical display supports peak search, marker functions, and comparative analysis. Its rugged construction and user-friendly interface allow deployment in both laboratory and field environments. The unit supports USB and GPIB connectivity for seamless data transfer and remote control. Designed for the evaluation of optical amplifiers, filters, and sources, it ensures accurate characterization of wavelength-dependent components in dense WDM systems.
Aumictech – Yokogawa AQ6360 Optical Spectrum Analyzer 1200–1650 nm Specifications
Optical Performance
Wavelength range: 1200 to 1650 nm
Span: 0.5 nm to full range, and 0 nm
Wavelength accuracy: ±0.02 nm (C-band, 1520–1580 nm), ±0.04 nm (1580–1620 nm), ±0.1 nm across full range
Wavelength linearity: ±0.01 nm (1520–1580 nm)
Wavelength repeatability: ±0.005 nm
Resolution bandwidth settings: 0.02, 0.05, 0.1, 0.2, 0.5, 1.0, 2.0 nm
Resolution accuracy: ±5 % (≥0.05 nm RBW, with resolution correction ON)
Minimum sampling resolution: 0.001 nm
Level / Power Measurement
Sensitivity modes: NORM_AUTO, NORMAL, MID, HIGH1, HIGH2, HIGH3
Sensitivity (typical): −80 dBm (1300–1600 nm, HIGH3 mode)
Maximum safe input power: +20 dBm
Level accuracy: ±0.3 dB (1310/1550 nm, −20 dBm input, MID/HIGH modes)
Level linearity: ±0.05 dB (−50 to +10 dBm)
Flatness: ±0.1 dB (1520–1580 nm), ±0.2 dB (1580–1620 nm)
Polarization dependence: ≤ ±0.05 dB
Dynamic Range & Stray Light
Dynamic range (typical):
• 70 dB at ±0.4 nm (0.05 nm RBW)
• 55 dB at ±0.2 nm (0.02 nm RBW)Stray light suppression ratio: ~75 dB typical
Overall level range: +20 to −80 dBm
Sweep & Speed
Sweep time (100 nm span, typical):
• FAST mode: ~0.2–0.5 s
• NORMAL mode: ~1–2 s
• HIGH SENS mode: 20–60 sSweep modes: single, repeat, and continuous
High dynamic mode: available to reduce stray light and improve close-in dynamic range
Display, Storage & Interfaces
Display: 10.4-inch color LCD, 800×600 resolution
Trace memory: up to 64 traces
Trace functions: subtraction, Max/Min hold, overlay comparison
Interfaces: GP-IB, RS-232C, Ethernet, USB, trigger in/out, VGA output
Data export formats: CSV, binary, BMP, TIFF
Printer option: built-in thermal printer available
Physical & Power
Dimensions (W×H×D): approx. 426 × 221 × 459 mm
Weight: approx. 19 kg (without printer)
Power supply: 100–240 VAC, 50/60 Hz, ~150 VA
Operating environment:
• Guaranteed performance: +18 to +28 °C
• Operating: +5 to +35 °C
• Storage: −10 to +50 °C
• Humidity: ≤80 % RH, non-condensingWarm-up time: ~1 hour
Comparison: AQ6360 vs AQ6370 Series
Wavelength coverage: AQ6360 spans 1200–1650 nm, while AQ6370C/D extend wider (600–1700 nm).
Resolution & accuracy: Both offer 0.02 nm resolution with similar ±0.02 nm accuracy in C-band; AQ6370D achieves ±0.01 nm in high-performance versions.
Dynamic range: AQ6360 provides ~70 dB close-in dynamic range; AQ6370D can exceed 75–80 dB depending on settings.
Positioning: AQ6360 is well-suited for telecom and fiber R&D focused on the 1200–1650 nm window, while AQ6370 series covers both telecom and broader R&D applications.
Why Choose Aumictech for the Yokogawa AQ6360?
The Yokogawa AQ6360 Optical Spectrum Analyzer is a reliable solution for telecom system testing, fiber optic development, and optical device characterization in the 1200–1650 nm band. Its balance of accuracy, sensitivity, and fast sweep speeds makes it a strong performer in both R&D and production environments.
At Aumictech, we provide:
NIST-traceable calibration for guaranteed wavelength and power accuracy
Professional repair and servicing to maximize instrument life
Expert guidance on whether the AQ6360 or AQ6370 series best fits your application needs


















Reviews
There are no reviews yet.