The Yokogawa AQ6370B is an optical spectrum analyzer that covers a wavelength range from 600 nm to 1700 nm. It offers high wavelength resolution and accuracy, making it suitable for evaluating optical transceivers, passive components, and fiber optic cables. Its built-in analysis functions and user-friendly interface enhance productivity and simplify complex measurements.
Aumictech – Yokogawa AQ6370B Optical Spectrum Analyzer 600–1700 nm Specifications
Optical Performance
Wavelength range: 600 to 1700 nm
Span: 0.5 nm to full range, and 0 nm
Wavelength accuracy: ±0.02 nm (1520–1580 nm), ±0.04 nm (1580–1620 nm), ±0.1 nm across the full range
Wavelength linearity: ±0.01 nm (1520–1580 nm)
Wavelength repeatability: ±0.005 nm
Resolution bandwidth settings: 0.02, 0.05, 0.1, 0.2, 0.5, 1.0, 2.0 nm
Resolution accuracy: ±5 % (≥0.05 nm RBW, resolution correction ON)
Minimum sampling resolution: 0.001 nm
Level / Power Measurement
Sensitivity modes: NORM_AUTO, NORMAL, MID, HIGH1, HIGH2, HIGH3
Sensitivity (typical): down to −85 dBm (1300–1600 nm, HIGH3 mode)
Maximum safe input power: +20 dBm
Level accuracy: ±0.4 dB (1310/1550 nm, −20 dBm input, MID/HIGH modes)
Level linearity: ±0.05 dB (−50 to +10 dBm)
Flatness: ±0.2 dB across 1520–1620 nm
Polarization dependence: ≤ ±0.05 dB
Dynamic Range & Stray Light
Dynamic range (typical):
• ~70 dB at ±1 nm (0.05 nm RBW)
• ~60 dB at ±0.4 nm (0.05 nm RBW)
• ~45 dB at ±0.2 nm (0.02 nm RBW)Stray light suppression ratio: ~70–73 dB typical
Overall level range: +20 to −85 dBm
Sweep & Speed
Sweep time (100 nm span, typical):
• FAST mode: ~0.5 s
• NORMAL mode: ~1–2 s
• HIGH SENS mode: 30–60 sSweep modes: single, repeat, and continuous
Display, Storage & Interfaces
Display: 10.4-inch color LCD, 800×600 resolution
Trace memory: up to 32–64 traces depending on configuration
Trace functions: subtraction, Max/Min hold, overlay comparison
Interfaces: GP-IB, RS-232C, Ethernet, USB, trigger in/out, VGA output
Data export formats: CSV, binary, BMP, TIFF
Printer option: built-in thermal printer available
Physical & Power
Dimensions (W×H×D): approx. 426 × 221 × 459 mm
Weight: approx. 19 kg
Power supply: 100–240 VAC, 50/60 Hz, ~150 VA
Operating environment:
• Guaranteed performance: +18 to +28 °C
• Operating: +5 to +35 °C
• Storage: −10 to +50 °C
• Humidity: ≤80 % RH, non-condensingWarm-up time: ~1 hour
Comparison: AQ6370B vs AQ6370 / AQ6370C / AQ6370D
AQ6370 (original): Strong baseline analyzer with wide coverage but less efficient in dynamic range.
AQ6370B: Improved user interface, enhanced stability, and better polarization dependence over the original AQ6370.
AQ6370C: Advanced version with better wavelength accuracy and additional high-performance (-20) option.
AQ6370D: Latest in the series, with higher accuracy (±0.01 nm in C-band for high-performance model), faster sweep speeds, and superior close-in dynamic range.
Why Choose Aumictech for the Yokogawa AQ6370B?
The Yokogawa AQ6370B Optical Spectrum Analyzer provides reliable performance across the 600–1700 nm wavelength range, making it ideal for telecom R&D, optical device testing, and WDM system evaluation. Its balance of speed, sensitivity, and resolution made it a workhorse during its release and it remains a capable legacy instrument.
At Aumictech, we provide:
NIST-traceable calibration for guaranteed measurement accuracy
Repair and service support to extend instrument life
Upgrade guidance for customers considering transitions to AQ6370C or AQ6370D models