The Yokogawa WT333E is a precision digital power meter engineered for AC and DC power measurements across R&D, production testing, quality assurance, and device evaluation environments. It delivers basic accuracy of ±(0.1% of reading + 0.05% of effective value measuring range) or ±0.15% of reading at 50/60 Hz, with simultaneous measurement of RMS, MEAN, and DC voltage alongside harmonic components and THD. The instrument supports multiple wiring configurations—single-phase three-wire, three-phase three-wire, three-phase four-wire, and three-voltage three-current—making it adaptable to diverse power distribution topologies.
– Technical Specifications
Measurement Range & Accuracy
• Voltage: 6 ranges from 15 V to 600 V RMS; up to 1200 V RMS with modified range (1000 V RMS maximum input)
• Current: 6 ranges from 0.5 A to 20 A direct input; down to 50 µA with specified accuracy; 5 mA minimum range with 100 µW resolution
• Frequency bandwidth: DC, 0.1 Hz to 100 kHz for both voltage and current
• Sampling rate: 100 kS/s
• Crest factor: Measurable up to 300× in all ranges
• Data update rate: 100/250/500 ms/1/2/5/10/20 s, Auto selectable
External Sensor Inputs
• EX1 channel: 2.5 V, 5 V, 10 V ranges
• EX2 channel: 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V ranges
Harmonics & Integration
• Harmonic measurement: Up to 50th order (requires /G5 option)
• THD calculation: Included with harmonics analysis; PLL synchronization supported
• Integration time: Up to 10,000 hours with measurements for energy (Wh) and current (Ah); automatic range setting in integration mode
– Key Features
• ±0.15% basic accuracy at fundamental frequency
• Simultaneous multi-parameter acquisition
• High-speed automatic ranging during integration
• Selectable data update rates from 100 ms to 20 s
– Typical Applications
• Power supply characterization and efficiency testing
• Motor and drive evaluation
• Home appliance energy auditing
• Automotive electronics power profiling
• Manufacturing quality control and acceptance testing
– Compatibility & Integration
• USB (standard)
• GPIB/IEEE-488 or RS-232-C selectable (-C2 option)
• Ethernet with Modbus/TCP (-C7 option)
• WTViewerE free PC application for parameterization, control, and data acquisition
• WTViewerFreePlus for monitoring and logging up to 200 parameters
















Reviews