The Fluke 724 Temperature Calibrator simultaneously measures and sources temperature signals across RTD, thermocouple, voltage, current, and resistance parameters. Designed for field and laboratory temperature sensor verification, it delivers calibration-grade accuracy with straightforward operation and dual-line display output. The instrument supports 10–12 thermocouple types (J, K, T, E, L, N, U, B, R, S, XK, BP) and 7 RTD variants (Ni120, Pt100 385/393/JIS/3926, Pt200, Pt500, Pt1000), plus DC voltage, current, and resistance measurement and sourcing functions.
– Technical Specifications
Temperature Measurement & Sourcing
• Thermocouple types: J, K, T, E, L, N, U, B, R, S, XK, BP
• RTD types: Ni120; Pt100 (385, 393, JIS, 3926); Pt200 (385); Pt500 (385); Pt1000 (385)
• RTD accuracy: 0.2 °C to 0.33 °C (range dependent)
• Thermocouple accuracy: 0.7 °C to 2.2 °C (range and type dependent)
Electrical Measurement & Sourcing
• Voltage measurement: 0–30 V DC; voltage sourcing: 0–10 V DC
• Millivolt measurement/sourcing: 0–100 mV DC; -10.00 mV to +75.00 mV
• Current measurement: 0–24 mA DC
• Resistance measurement: 0–3200 Ω; resistance sourcing: 15–3200 Ω
• Voltage DC measurement accuracy: 0.02% Rdg + 2 LSD; 0.025% + 1 count (TC connector)
• Voltage DC source accuracy: 0.02% Rdg + 2 LSD; 0.025%
• Current DC accuracy: 0.02% Rdg + 2 LSD
• Resistance accuracy: 0.1 Ω to 1.5%; 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
• Resistance source accuracy: 0.15 Ω to 1 Ω
• Temperature coefficient: ±0.005% of range per °C
Loop Power
• 24 V DC supply output; 10% accuracy
– Key Features
• Simultaneous input and output display
• No-menu control interface
• 24 V DC loop power for transmitter testing
• 4-wire, 2-wire, and 3-wire RTD measurement modes
• Compact form factor: 96 × 200 × 47 mm
– Typical Applications
• RTD and thermocouple sensor calibration and verification
• Temperature transmitter loop testing
• Field instrument commissioning and maintenance
• Industrial process temperature validation
– Compatibility & Integration
The 724 measures and sources across industry-standard temperature sensor types and accepts voltage, current, and resistance signals for general-purpose instrumentation verification.



























Reviews