The HP/Agilent 6811A is a 375 VA AC power source and analyzer engineered for precise generation and measurement of AC power parameters across development, manufacturing, and research environments. It functions as both a programmable power source and real-time analyzer, enabling waveform synthesis, power line disturbance simulation, and detailed harmonic characterization in a single benchtop instrument.
– Technical Specifications
Output Power and Voltage
• 375 VA maximum power output
• 0–300 V RMS AC output; ±425 V DC capability
• 0–3.25 A RMS AC; 0–2.5 A DC maximum
• DC to 1 kHz operating frequency range (45 Hz to 1 kHz at full rating)
Measurement Accuracy
• RMS voltage (45–100 Hz): 0.15% + 0.3 V
• DC voltage: 0.1% + 0.5 V
• RMS current (45–100 Hz, high range): 0.05% + 10 mA
• Power (VA, 45–100 Hz): 0.1% + 1.5 VA + (0.05% + 0.05%/kHz + 250 mV)
• Power (watts, 45–100 Hz): 0.1% + 0.3 W + (0.05% + 0.05%/kHz + 250 mV)
• Frequency: 0.01% + 0.01 Hz
• Power factor resolution: 0.01
• 16-bit measurement resolution
Waveform and Distortion Capability
• Output types: sine, square, clipped sine, and up to 12 user-defined arbitrary waveforms
• Harmonic analysis to the 50th harmonic (≤250 Hz output frequencies) with magnitude and phase
• THD measurement for voltage and current
• Maximum output THD: 0.25% at 50/60 Hz
• Voltage ripple and noise: −60 dB
Power Line Disturbance Simulation
Generates sag, surge, dropout, and clipping events with programmable amplitude, frequency, phase, slew rate, and transient modes (step, pulse, list).
– Key Features
• Programmable voltage, current limit, frequency, phase, and distortion parameters
• Real-time measurement of RMS voltage, RMS current, peak current, frequency, phase, real power, reactive power, apparent power, and power factor
• Line regulation: 0.1%; load regulation: 0.5%
• Universal AC input: 100–240 V, 1 Ph
– Typical Applications
Product development and validation; manufacturing test; harmonic compliance assessment; transient response characterization; power factor measurement and analysis.
– Compatibility & Integration
HP-IB (IEEE 488) and RS-232 interfaces with SCPI programming.