The Fluke 6100A Electrical Power Standard is a precision calibration instrument for verification of electrical power measurement devices across design engineering, manufacturing, service, and standards laboratory environments. This configurable system generates pure sinusoidal waveforms and complex signals including harmonics, interharmonics, fluctuating harmonics, flicker, dips, and swells to simulate real-world power conditions. Single-phase or multi-phase operation (up to four phases) is supported through addition of 6101A Auxiliary units, enabling compliance validation against IEC 61000-4-15 and IEC 61000-4-7 standards.
– Technical Specifications
Output Characteristics:
• Voltage output: up to 1000 V AC
• Current output: up to 21 A AC (optionally up to 80 A with booster)
• Voltage/current amplitude resolution: 6 digits
• Amplitude accuracy: approaching 100 ppm (0.01%)
• Frequency range: 16 Hz to 450 Hz
• Frequency accuracy: 50 ppm
• Frequency resolution: 0.1 Hz
• Phase angle adjustment: −180° to +180°
• Phase angle resolution: 0.001° or 10 micro-radians
• Phase accuracy: within 3 millidegrees
• Phase stability (short term): within 200 micro-degrees
• Multi-phase phase accuracy: within 5 millidegrees between phases
• Power output (voltage terminals): up to 50 VA
• Compliance voltage (current output): up to 14 V
• Auxiliary voltage (current output): up to 10 V
• Warm-up time to full accuracy: 1 hour
• Settling time: less than 1.4 seconds
Signal Generation:
• Harmonics: up to 100th harmonic simultaneously
• Maximum harmonic frequency: 6 kHz
• Maximum individual harmonic level: up to 30% of fundamental
• Harmonic setting methods: % RMS, % fundamental, dB down from fundamental, absolute value
• DC offset: programmable
• Interharmonics: supported
– Key Features
• Scalable architecture from single-phase to four-phase configurations
• Six-digit resolution for precision amplitude control
• Complex waveform simulation for standards compliance testing
• Long compliance voltage ensures delivery across extended cable runs
– Typical Applications
• Power meter calibration and verification
• Harmonic distortion testing
• Flicker and voltage event simulation
• Multi-phase power system validation
– Compatibility & Integration
Expansion via 6101A Auxiliary units enables multi-phase operation without requiring separate primary instruments.



























Reviews
There are no reviews yet.