The Ando AQ6317B Optical Spectrum Analyzer is designed for precise optical spectrum analysis. With a wide wavelength range and high resolution, it enables detailed analysis of optical signals. The recent calibration ensures that the instrument meets specifications and provides dependable measurements, making it suitable for research, development, and manufacturing environments where accuracy is critical.
Aumictech – Ando AQ6317B Optical Spectrum Analyzer 600–1750 nm Specifications
Optical Performance
Wavelength range: 600 to 1750 nm
Wavelength accuracy:
• ±0.02 nm (1520–1580 nm, calibrated with internal reference)
• ±0.04 nm (1580–1620 nm)
• ±0.5 nm across full 600–1750 nm spanWavelength linearity: ±0.01 nm (1520–1580 nm), ±0.02 nm (1580–1620 nm)
Wavelength repeatability: ±0.005 nm (1 min)
Resolution bandwidth settings: 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1.0, 2.0 nm
Resolution accuracy: ±5 % (for ≥0.05 nm settings, 1300–1650 nm, resolution correction ON)
Minimum sampling resolution: 0.001 nm
Level / Power Measurement
Level range:
• −90 to +20 dBm (1200–1650 nm, HIGH3 sensitivity)
• −80 to +20 dBm (1000–1200 nm, HIGH3 sensitivity)
• −60 to +20 dBm (600–1000 nm, HIGH3 sensitivity)Level accuracy: ±0.3 dB (1310/1550 nm, −30 dBm input, HIGH1–3)
Level linearity: ±0.05 dB (−50 to +10 dBm, HIGH1–3)
Flatness: ±0.1 dB (1520–1580 nm), ±0.2 dB (1580–1620 nm)
Polarization dependence: ≤ ±0.05 dB
Dynamic Range & Stray Light
Dynamic range (typical):
• ~70 dB (±0.4 nm at 0.01 nm resolution)
• ~60 dB (±0.2 nm at 0.01 nm resolution)Stray light suppression ratio: ~75 dB typical
Level range: +20 to −90 dBm
Sweep & Speed
Sweep time (100 nm span, typical): ~0.5 s in FAST mode, ~2 s in NORMAL mode, ~30–60 s in HIGH SENS mode
Sweep modes: single, repeat, and continuous
Display, Storage & Interfaces
Display: 9.4-inch color LCD, 640×480 resolution
Memory: up to 32 traces and 20 user programs storable internally
Trace functions: subtraction, Max/Min hold, overlay comparison
Interfaces: GP-IB (2 ports), RS-232C, VGA output, trigger in/out, analog output
Printer support: external GPIB printer
Physical & Power
Dimensions (W×H×D): approx. 425 × 222 × 450 mm
Weight: approx. 30 kg
Power supply: AC 100–240 V, 50/60 Hz
Operating environment: +5 to +40 °C; ≤80 % RH (non-condensing)
Warm-up time: ~1 hour
Comparison: AQ6317B vs Similar Models
Versus AQ6317 (non-B):
The AQ6317B improves wavelength accuracy (±0.02 nm in C-band vs ~±0.05 nm in older versions), adds faster sweep times, and expands analysis functions such as optical filter testing and multi-channel noise figure evaluation.Versus AQ6370C/D:
Yokogawa AQ6370 series analyzers are newer successors with enhanced resolution (down to 0.02 nm standard, ±0.01 nm accuracy in high-performance models) and higher dynamic range in some configurations. However, the AQ6317B remains a reliable and cost-effective solution with strong performance across the 600–1750 nm range.
Why Choose Aumictech for the Ando AQ6317B?
The Ando AQ6317B Optical Spectrum Analyzer is a trusted instrument for telecom, photonics, optical component testing, and WDM system analysis. Its high resolution, strong sensitivity, and robust dynamic range make it an enduring workhorse in R&D and production labs.
At Aumictech, we provide:
NIST-traceable calibration for precise wavelength and power accuracy
Repair and service support to extend instrument life
Upgrade guidance to help customers evaluate when to keep legacy analyzers like the AQ6317B or transition to newer models



















Reviews