The Keysight 6032A is a 1064 W autoranging DC power supply engineered for test systems demanding high power and precise voltage/current control. Single-output design with continuously variable 0–60 Vdc and 0–50 A capability delivers maximum power across multiple operating points: 60 V at 17.5 A, 40 V at 30 A, and 20 V at 50 A. Programming accuracy reaches ±0.035% + 40 mV on voltage and ±0.2% + 85 mA on current. Load regulation holds voltage to 5 mV and current to 10 mA. Ripple and noise stay below 8 mV rms and 40 mV peak-to-peak across the 20 Hz to 20 MHz band. The unit integrates GPIB (IEEE 488) control with SCPI command support, enabling integration of up to 16 supplies through a single GPIB address via serial link. Remote sensing accommodates up to 2 V lead drop while maintaining regulation performance. Internal fan cooling, Safety Class 1 construction, and compact dimensions (133 mm H × 426 mm W × 504 mm D) suit laboratory, system integration, and field engineering applications. Input: 104–127 Vac, 48–63 Hz.
– Technical Specifications
• Output voltage: 0–60 Vdc
• Output current: 0–50 A
• Power output: 1064 W
• Autoranging capability: 60 V/17.5 A, 40 V/30 A, 20 V/50 A
• Programming accuracy (voltage): ±0.035% + 40 mV
• Programming accuracy (current): ±0.2% + 85 mA
• Readback accuracy (voltage): ±0.08% + 20 mV
• Readback accuracy (current): ±0.36% + 35 mA
• Load regulation (voltage): 5 mV; (current): 10 mA
• Line regulation (voltage): 0.01% + 5 mV
• Ripple and noise (20 Hz–20 MHz): 8 mV rms / 40 mV peak-to-peak (voltage); 25 mA rms (current)
• Transient response (10% step): 2 ms @ 150 mV
• Remote sensing: up to 2 V per lead
• Input: 104–127 Vac, 48–63 Hz
– Key Features
• GPIB (IEEE 488) interface with SCPI command support
• Serial link capability for multi-supply configurations (up to 16 units per address)
• Remote sensing for extended lead lengths
• Internal fan cooling
• Safety Class 1, protective earth terminal
– Typical Applications
• Automated test equipment and system integration
• High-current circuit validation and stress testing
• Engineering bench applications requiring variable power
– Compatibility & Integration
GPIB connectivity and VXI Plug&Play driver support enable integration into existing automated test frameworks. Multiple units chain via serial link for scalable multi-supply systems.





















Reviews
There are no reviews yet.