The Agilent/Keysight 6034A is a 200 W autoranging DC power supply delivering 0–60 Vdc at 0–10 Adc for laboratory and system test applications. FET switching topology enables full power availability across the entire voltage–current operating envelope without manual range switching. The unit provides HP-IB (IEEE-488) programmability for voltage, current, and overvoltage protection settings, plus integrated measurement capability for closed-loop control and real-time monitoring of output conditions.
– Technical Specifications
• Output Range: 0–60 Vdc / 0–10 Adc (200 W maximum)
• Programming Resolution: 15 mV voltage (12-bit); 2.5 mA current (12-bit)
• Programming Accuracy (25 ±5°C): ±0.07% ±28 mV (CV mode); ±0.085% ±12.5 mA (CC mode)
• Output Impedance: Typically 1 mΩ series with 30 µH; <1 Ω below 1 MHz
• DC Output Isolation: ±240 Vdc from ground
• Overvoltage Protection: Local range 1.7–64.5 V (0.2 V resolution); remote range 2.0–64.5 V (0.25 V resolution); ±0.7 V accuracy
• AC Input: 104–127 Vac (standard)
• Operating Environment: ≤95% RH, up to 2000 m altitude, indoor Category II, Pollution Degree 2
• Cooling: Internal fan with thermal management
– Key Features
• Autoranging topology eliminates manual band switching for flexible voltage–current combinations
• 12-bit programming resolution enables fine control across full output range
• HP-IB interface supports remote voltage and current programming plus OVP adjustment
• HP-IB measurement feedback allows real-time monitoring of actual output voltage and current
• Programmable overvoltage protection with dual adjustment modes (local and remote)
• Built-in overload and short-circuit protection
• Service Request (SRQ) capability for fault notification
– Typical Applications
• Component and subsystem bench testing requiring stable 60 V/10 A output
• Automated test systems with closed-loop voltage or current control
• Development and validation of power-sensitive circuits and modules
• Laboratory evaluation of loads across the full 200 W operating range
– Compatibility & Integration
HP-IB (IEEE-488) connectivity enables integration into instrumentation systems and ATE environments. Output impedance characteristics support both steady-state and dynamic load conditions with minimal ripple across standard test frequencies.


















Reviews