The Agilent/Keysight 83050A is a microwave system amplifier engineered for integration into automated test equipment and RF/microwave measurement systems. Operating across 2 GHz to 50 GHz, it delivers gain exceeding 21 dB across most of its bandwidth, with 30 dB typical performance in the 1–2 GHz band. The amplifier consolidates the function of multiple narrow-bandwidth units into a single platform, eliminating the need for complex crossover networks and reducing bias supply complexity.
– Technical Specifications
• Frequency Range: 2 GHz to 50 GHz
• Gain: 21 dB minimum; 30 dB typical (1–2 GHz); 27 dB minimum (20–26.5 GHz)
• Output Power (P1dB): +20 dBm (2–40 GHz); +12 dBm (to 45 GHz); +30 dBm typical (1–2 GHz)
• Noise Figure: Less than 6 dB typical (2–26.5 GHz); less than 10 dB typical (10–50 GHz)
• Harmonic Distortion: –22 dBc typical (1–2 GHz at P1dB); –20 dBc typical (2–18 GHz at P1dB)
• Output SWR: 2.8 max (2–18 GHz); 2.2 max (18–50 GHz)
• Impedance: 50 Ω typical
• Rise Time: 250 ps typical
– Key Features
• RF connectors: 2.4 mm (female)
• DC detector output: BNC (female)
• Bias requirements: +12 V @ 830 mA, –12 V @ 50 mA
• Compatible with 87421A 25 W power supply
• Operating environment: Category II installation, Pollution Degree 2
• Temperature stabilization: 1 hour continuous operation; +20°C to +30°C for warranted specs
• Maximum humidity: 95% relative; altitude to 2000 meters
– Compatibility & Integration
Part of the 83000A Series microwave system amplifier family. Replaces multiple narrow-bandwidth amplifiers in integrated system designs.