The Agilent/Keysight 87510A Gain-Phase Analyzer is a precision instrument for measuring gain and phase characteristics of electronic components and systems from 100 kHz to 300 MHz. It delivers vector transmission measurements with 1 mHz frequency resolution and dynamic accuracy of ±0.05 dB and ±0.3°. The analyzer supports multiple measurement modes (A/R, A, R) and features automatic parameter-extraction commands for filter and resonator characterization.
– Technical Specifications
Signal Source
• Frequency range: 100 kHz to 300 MHz
• Frequency resolution: 1 mHz
• Output level range: −15 dBm to +5 dBm (standard); −45 dBm to +15 dBm with Option 010
• Output level accuracy: ±1 dB at 50 MHz, 0 dBm, 23°C ±5°C
• Frequency stability: ±5 × 10⁻⁷/day typical at 23°C ±5°C; ±2.5 × 10⁻⁹/8 h with Option 001
• Frequency accuracy: ±20 ppm at 23°C ±5°C; ±1 ppm with Option 001
Receiver
• Frequency range: 100 kHz to 300 MHz
• IF bandwidth: 20 Hz, 200 Hz, 1 kHz, 4 kHz, 8 kHz
• Maximum input level: 0 dBm
• Noise floor: −110 dBm (IFBW = 20 Hz)
• Dynamic accuracy: ±0.05 dB, ±0.3° at IFBW 20 Hz, −10 dBm (A and R channels)
Power Splitter (standard, except Option 004)
• Insertion loss: 6 dB nominal
• Tracking: 0.1 dB (100 kHz to 100 MHz); 0.2 dB (>100 MHz)
Input Connectors
• Standard: BNC, 50 Ω (two inputs: R, A)
• Option 003: Type-N (A channel), BNC (R channel)
• Option 004: Single BNC input (A channel only)
– Key Features
• High-speed vector transmission measurement: 0.25 ms per point at 8 kHz IFBW
• Automatic filter and resonator parameter extraction
• Selectable IF bandwidths for noise-sensitivity optimization
• SWEEP END output signal: 10 µs negative pulse, +5 V logic level, 100 mA maximum current
– Typical Applications
• Amplifier and filter characterization
• Audio system analysis
• Resonator production line testing
• Component gain-phase measurement in R&D and manufacturing
– Compatibility & Integration
Input impedance is 50 Ω nominal. Connector options accommodate standard BNC installations or high-frequency Type-N implementations. Maximum input damage level: 20 dBm, 50 Vdc.



























Reviews