The Agilent/Keysight E1413C is a 16-bit, 64-channel differential scanning ADC module housed in a single VXI C-size slot. It delivers up to 100,000 samples/second across configurable channel lists and integrates a 65,024-sample dual-ported FIFO buffer for sustained high-speed data transfer. The module accepts up to eight Signal Conditioning Plug-on (SCP) modules, enabling DC voltage, temperature, resistance, and strain measurements across input ranges from 4 mV to 16 V (expandable to 60 V with attenuator SCP). Autoranging gains of ×1, ×4, ×16, ×64, and ×256 provide measurement flexibility. Data output is IEEE 754 32-bit floating-point, either as engineering units or raw input voltage.
– Technical Specifications
• ADC Resolution: 16-bit
• Channel Capacity: 64 differential channels
• Maximum Scan Rate: 100,000 samples/second (divided by active channels in scan list)
• Data Buffer: 65,024 samples dual-ported FIFO; Current Value Table (CVT) for real-time per-channel monitoring
• Measurement Accuracy: 0.01% of reading
• Input Range (base): 4 mV to 16 V
• Input Range (with E1513A SCP): Up to 60 V
• Autoranging Gains: ×1, ×4, ×16, ×64, ×256
• Data Format: IEEE 754 32-bit floating-point
• Scan Lists: Four lists, each up to 1,024 channel assignments; List of Lists references up to 1,024 times
• Sample Timer: 10 µsec to 32,768 msec with 0.5 µsec resolution
• Scan Trigger Timer: 100 µsec to 6.5536 sec with 100 µsec resolution
– Key Features
• Register-based VXI architecture with single-slot C-size form factor
• Automatic self-calibration with on-board calibration sources
• Wiring offset compensation capability
• TTL trigger input from external hardware or other VXI modules
• Signal Conditioning Plug-on (SCP) modularity accommodates eight modules (eight channels per SCP typical)
– Typical Applications
• Multi-channel data acquisition in production test
• Precision DC voltage and resistance measurement
• Temperature and strain monitoring in environmental test chambers
• High-speed scanning across large channel counts
– Compatibility & Integration
The E1413C operates as a register-based VXI module within standard VXIbus mainframes. SCP conditioning options tailor input stages for specific measurement types, and the Current Value Table enables real-time monitoring alongside buffered data collection.