The Agilent/Keysight E1416A is a single-slot, C-size VXI message-based power meter module engineered for accurate average power measurements across RF and microwave frequencies. This single-channel instrument integrates directly into VXI-based test systems and accepts Agilent 8480 series thermocouple and diode power sensors, plus the 11722A sensor, enabling flexible power measurement from 100 kHz to 110 GHz.
– Technical Specifications
Module Architecture
• Device type: VXI message-based
• Form factor: Single-slot C-size module
• Power consumption: 12.20 W per slot; 0.20 liter/s airflow; 1.00 mm H₂O pressure drop
• Supply currents: +5 V @ 1.0 A; +24 V / -24 V @ 0.1 A each
Power Reference
• 1.00 mW calibration standard (50 MHz oscillator, factory set to ±0.7%, traceable to NIST)
Measurement Performance
• Frequency range: 100 kHz to 110 GHz (sensor dependent)
• Power range: -70 dBm to +44 dBm (100 pW to 25 W) (sensor dependent)
• Dynamic range: 50 dB in 10 dB steps
• Absolute accuracy: ±0.02 dB or ±0.5%
• Zero-set repeatability: ±0.5% of full scale on most sensitive range; divided by 10 for each higher range; ±1 count
• Resolution: 0.1, 0.01, or 0.001 dB (selectable)
• Averaging: 1 to 1024 readings (powers of 2)
– Key Features
• Measurement results in W, dBm (absolute), %, or dB (relative)
• Offset capability: ±99.99 dB
• Duty cycle mode: Displays peak power representation for rectangular pulses
• Sensor calibration: Stores up to 10 frequency-dependent cal factor tables
• Operating state memory: Saves and recalls 10 complete configurations
– Compatibility & Integration
Supported Sensors: Agilent 8480 series (thermocouple and diode types), 11722A
Functional Equivalence: Agilent 437B or 70100A
Cable Options: 5-ft (1.5 m) or 10-ft (3.0 m) power sensor cables (P/N 11730A or 11730B)
Available Options: Custom sensor cable lengths (Option 004); service manual P/N E1416-90002 (Option 915); calibration service (CAL)