The Agilent/Keysight E3615A is a 60 W single-output DC power supply engineered for benchtop test and measurement applications. It delivers 0–20 VDC at 0–3 ADC with dual digital meters for simultaneous voltage and current monitoring, plus independent 10-turn potentiometers for precise output adjustment. The unit operates in both constant voltage (CV) and constant current (CC) modes to accommodate diverse load conditions.
– Technical Specifications
Output Ratings
• Maximum power: 60 W
• Voltage range: 0 to 20 VDC
• Current range: 0 to 3 ADC
Regulation & Stability
• Load and line regulation (CV): < 0.01% + 2 mV
• Load regulation (CC): < 0.01% + 250 µA
• Line regulation (CC): < 0.01% + 250 µA
• Voltage ripple and noise: < 200 µVrms, < 1 mVp-p
• Output drift (8 hours after 30 min warm-up): < 0.1% + 5 mV (voltage), < 0.1% + 10 mA (current)
• Transient recovery (full to half load): < 50 µsec within 15 mV
Metering
• Accuracy: ±0.5% + 2 counts
• Digital displays for voltage and current
Protection & Isolation
• Adjustable overvoltage protection
• Output isolation to earth ground: ±240 Vdc maximum
• Either output terminal floats independently at up to 240 V from ground
Thermal & Environmental
• Cooling: Convection
• Operating temperature: 0 to 40 °C (full rated output); current derates 1% per °C from 40 °C to 55 °C
• AC input: 100 Vac ±10%, 47–63 Hz (standard); optional 230 Vac ±10% or 115 Vac ±10%
– Key Features
• Remote analog voltage programming: 0–10 V input maps to 0–20 V output with 0.5% linearity; inputs protected to ±40 V
• Remote sensing compensates for voltage drops in load leads
• Front and rear output terminals for flexible connectivity
• Three-conductor mains power cable required
– Typical Applications
Laboratory power sourcing, component characterization, device testing under CV and CC control, and systems requiring low-noise DC with remote programming and lead-drop compensation.
– Compatibility & Integration
Analog programming interface accepts 0–10 V control signals. Output floats independently, enabling series or parallel configurations within isolation limits. Either output terminal may be grounded.