The Agilent/Keysight E9325A is a peak and average power sensor engineered for precision power measurements across RF and microwave frequencies from 50 MHz to 18 GHz. This instrument delivers dual measurement modes—Normal Mode for peak, average, and time-gated analysis of modulated signals, and Average Only Mode optimized for low-level signal characterization. The sensor achieves measurement speeds up to 1000 corrected readings per second via GPIB in Fast mode, with settling times as low as 0.08 seconds in average-only configuration.
– Technical Specifications
• Frequency Range: 50 MHz to 18 GHz
• Measurement Modes: Normal (peak/average/time-gated); Average Only (optimized for low-level signals)
• Power Range—Average Only Mode: −65 dBm to +20 dBm
• Power Range—Normal Mode: −50 dBm to +20 dBm
• Maximum Power Handling: +23 dBm average; +30 dBm peak (<10 µsec)
• Video Bandwidth: 300 kHz (ideal for TDMA formats: NADC, GSM, EDGE)
• Measurement Speed: Up to 1000 corrected readings/second via GPIB (Fast mode); settling time 0.08 to 6.7 seconds depending on averages
• Linearity (Average Only): ±0.02 dB logarithmic; ±0.5% linear
• SWR: 1.12 (50 MHz–2 GHz); 1.16 (2–10 GHz); 1.23 (10–16 GHz)
• Connector: Precision Type-N (male)
• Dimensions: 150 mm L × 33 mm W × 30 mm H
• Weight: 0.2 kg net
– Key Features
• Dual-mode operation supports both peak detection and average power analysis
• EEPROM-stored calibration factors and temperature/bandwidth corrections
• Zero set capability: 5 nW (Normal mode), 0.17 nW (Average Only mode)
• Compact form factor enables field deployment and integration into test racks
• GPIB interface for automated measurement sequences
– Typical Applications
• Wireless communications power monitoring (GSM, EDGE, NADC)
• Radar and electronic warfare system characterization
• RF and microwave signal chain validation
• Time-gated power analysis in burst-mode systems
– Compatibility & Integration
Compatible with EPM-P Series (E4416A/E4417A) and P-Series (N1911A/N1912A) power meters for peak, average, and time-gated measurement modes. Also works with EPM Series meters for average power analysis. Requires E9288A, B, or C sensor cable (E9288B/C supplied in 3 m or 10 m lengths).