The Associated Research 5500DT is a HypotPLUS II AC Withstand Voltage Tester, designed for comprehensive electrical safety testing. This instrument is engineered to ensure product safety and compliance with various international regulatory standards, including UL, CSA, VDE, IEC, and TÜV, as well as European Norms such as the Low Voltage Directive 73/23/EEC. Its robust design and advanced features make it suitable for both development and production line testing environments.
### Technical Specifications
* **Model:** 5500DT (AC Only Hipot with Ground Continuity Check)
* **Input Voltage:** 115/230 VAC
* **Output Voltage (AC):** 0-5000 VAC
* Adjustable in 10-volt steps
* **Output Current (AC):** 20 mA
* High Trip Limit: 0.05-20.00 mA AC
* Low Trip Limit: 0.00-19.99 mA AC
* Resolution down to 10 microamps
* **AC Output Waveform:** Sine wave with distortion ≤ 1%
* **Dwell Timer:** 0/0.5-999.9 seconds
* In 0.1-second increments or continuous
* Accuracy: ± (0.1% of reading + 0.05 seconds)
* **Ramp Time:** 0/0.5-999.9 seconds
* **Ground Continuity Test:**
* Trip current: 0.1 Amp
* Continuity Trip: 1.1 Ohm maximum
* **Arc Detection:** Digitally set
* **Failure Detection System:** Includes continuity, LO, HI, Arc, Short, and Breakdown
* **Test Programs:** Storage of up to 10 different test programs
* No load setup of trip current and output voltage
### Electrical
* **Hipot Testing Capabilities:** AC withstand voltage testing
* **Leakage Current Measurement:** Resolution down to 10 microamps
* **Voltage Control:** Precise voltage control and measurement
* **Real Current Option:** Available (includes 3kV AC @ 35mA)
### Interfaces
* **PLC Remote Inputs and Outputs:**
* Enhanced PLC & Ten Volt analog signal for automated applications
* Up to three program memories can be accessed through the remote control (PLC) port
* **Remote Interlock Function:** Option available
* **Remote Memory Access:**
* **Test Lead Connections:** High voltage output terminal, return terminal, continuity test terminal
### Physical
* **Dimensions:** 17″ x 3.5″ x 16.5″ (W x H x D)
* **Weight:** 24.00 lbs, 25 lbs
* **Enclosure:** Bench or rack mountable (2U height) with tilt-up front feet
* **Display:** High resolution LCD displays setup parameters and test results
* **Controls:** Microprocessor digital design with software-friendly menuing, tamper-proof front panel controls, single function dedicated keys for test setup, numeric keys for data entry and test program select
* **Indicators:** Pass lamp, Fail lamp, Flashing red LED when high voltage is activated
### Compatibility and Standards
* **Compliance:** UL, CSA, VDE, IEC, TÜV, and other safety agencies
* **Directives:** Low Voltage Directive 73/23/EEC, EMC Directive 89/336/EEC
* **Standards:** Conforms to EN 61010-1: 1993 + A2, EN 61010-031: 1994, EN 50082-1: 1992, EN 50081-1: 1992, IEC 801-2: 1984, EN 55022: 1994, IEC 801-3: 1984, EN 60555-2: 1987, IEC 801-4: 1988
* **Interconnection:** Can be interconnected with HYAMP II ground bond tester to form a semi-automated test system
### Calibration and Maintenance
* **Calibration:** Traceable to National Institute of Standards and Technology (NIST)
* Controlled by software, adjustments made through front panel keypad in a restricted access calibration mode
* Calibration information stored in non-volatile memory
* Calibration services available, including pre and post calibration data and 17025 accreditation
* **Service Interval:** Recommended annual calibration and inspection of safety-related components
Why Choose Aumictech for Associated Research 5500DT?
At Aumictech, we specialize in supplying high-end electrical safety testing equipment. Whether you need reliable hipot testing, accurate insulation resistance measurement, or robust ground bond verification, our team ensures the Associated Research 5500DT delivers maximum performance and compliance for your application.