The Boonton 4300-4321 is a configurable RF power meter system designed for high-accuracy, high-resolution, and high-speed power measurements. It is engineered to accommodate various applications within component testing, system integration, and automated test equipment (ATE) environments. The modular design allows for a flexible configuration, enabling it to adapt to evolving measurement requirements.
### Technical Specifications
* **Model:** Boonton 4300 RF Power Meter Chassis (The 4321 is a specific calibrator module for this chassis).
* **Measurement Capability:** Up to six simultaneous, independent power measurements.
* **Frequency Range:** 10 kHz to 100 GHz, sensor-dependent. Some sources indicate up to 110 GHz.
* **Power Range:** -70 dBm to +44 dBm, sensor-dependent. Other sources list -75 dBm to +50 dBm or -70 dBm to +37 dBm.
* **Dynamic Range:**
* Up to 90 dB with diode sensors.
* Up to 50 dB with thermocouple sensors.
* **Measurement Throughput:** As fast as 0.01 seconds.
* **Settling Time:**
* Diode Sensors: < 200 ms for a 10 dB power step, hold range, filter set to 20 ms.
* Thermocouple Sensors: < 450 ms for a 10 dB power step, hold range, filter set to 20 ms.
* **Readout Speed:**
* Free run access time typically 30 ms per channel.
* Up to 30 readings per second over GPIB for a single channel.
* Up to 100 readings per second over GPIB for a six-channel meter.
* 20 milliseconds plus 5 milliseconds per channel over the IEEE bus.### Modules and ConfigurationThe Boonton 4300 is a chassis-based system that accommodates various input and calibrator modules:* **4300 Chassis:** The main unit with six module slots for input channels and one calibrator slot.
* **4311 Universal Channel Module:** Designed to operate with all Boonton Sensors, providing sensitivity and flexibility for up to 90 dB dynamic range with a single sensor.
* **4312 Thermocouple Channel Module:** Specifically designed for thermocouple sensors, expanding the dynamic range to 50 dB.
* **4321 30 MHz Programmable Calibrator Module:** (Front or Rear Output) Significantly improves basic measurement accuracy by allowing sensor linearity shaping to reduce errors to 0.003 dB.
* **4322 30 MHz Programmable Calibrator Module:** (Rear Output) Functionally similar to the 4321.
* **4323 50 MHz (0 dBm) Power Reference Module:** (Front Output) Provides a fixed 50 MHz, 0 dBm signal for applications requiring 50 MHz traceability. Used when the programmable calibrator's accuracy is not essential.
* **4324 50 MHz (0 dBm) Power Reference Module:** (Rear Output) Functionally similar to the 4323.
* **Optional Accessory Cables:** Available in lengths of 10 ft, 20 ft, 50 ft, and 100 ft, allowing sensors to be located up to 100 feet from the instrument.### Electrical Characteristics* **Power Consumption:** 40 VA.
* **Input Voltage:** 100, 120, 220, 240 V ±10%, 50 - 400 Hz.### Interfaces* **IEEE-488 Bus:** Complies with IEEE-488 standards, implementing SH1, AH1, T6, L4, SR1, RL1, DC1, and DT1. The open SEAbus Plus Protocol enables communication with host PCs and other systems.### Physical Characteristics* **Dimensions:** 3.485 in (8.8 cm) high, 17.24 in (43.8 cm) wide, and 17.75 in (45.0 cm) deep.
* **Weight:** 21 lbs (9.5 kg) with 6 channels and a programmable calibrator.
* **Environmental Compliance:** Meets MIL-STD-28800D Type II, Class 5, Style E and F equipment standards.
* **Operating Temperature:** 0°C to 55°C.
* **Storage Temperature:** -55°C to +75°C.### Features and Functionality* **Channel Math:** Allows mathematical operations between all six channels.
* **Zeroing:** Each range is independently zeroed via front panel or bus.
* **Limits:** Front panel or bus selectable between +100 dB and -100 dB, independently for each channel.
* **Filtering:** Selected by entering filter time in ms (20 ms to 20 s), accomplished in 20 ms increments (e.g., 2000 ms selects 100-point, pipeline averaging).
* **Ranging:** Autoranging, set-ranging, and hold on range; each channel operates independently. Hold on range offers 10% headroom and 25 dB dynamic range per range (0.2 dB resolution at -25 dB point).
* **Sensor Compatibility:** Compatible with all Boonton diode and thermocouple sensors. A wide selection of sensors are available to cover various frequency and power ranges.
* **Duty Cycle Entry:** Supported for pulse power measurements.
* **Upgradability:** Advanced system architecture supports simple upgrades of the meter's on-board operating software.### Compatibility* **Power Sensors:** Compatible with all Boonton diode and thermocouple sensors. Specific sensor models mentioned include 51071, 51072, 51075, 51077, 51079, 51100, 51101, 51102, 51200, 51201, 51300, 51301, 51085, 51086, 51087, 51011, 51012, 51013, 51015, and 51033.
* **ATE Systems:** MATE compatible. The SEAbus Plus Protocol enables integration with other systems.Why Choose Aumictech for the Boonton 4300-4321?
At Aumictech, we specialize in supplying high-end RF power measurement equipment. Whether you need precise, multi-channel power analysis, advanced calibration capabilities, or seamless integration into automated test systems, our team ensures the Boonton 4300-4321 delivers maximum performance for your application.