The Fluke 5502A Multi-Product Calibrator is a multifunction source and measurement instrument for calibrating electrical test equipment across calibration laboratories and service centers. It generates DC and AC voltage, current, resistance, and capacitance, while simultaneously sourcing voltage and current to simulate DC and AC power conditions. The unit sources thermocouple and RTD temperature signals, supports optional oscilloscope calibration to 600 MHz, and extends current capability through external amplifiers and coils. RS-232 and IEEE-488 interfaces enable remote control and integration with MET/CAL® Plus Calibration Management Software for automated workflows.
– Technical Specifications
Voltage and Current Output:
• DC Voltage: 0 to ±1020 V
• DC Current: 0 to ±20.5 A
• AC Voltage: 1 mV to 1020 V, 10 Hz to 500 kHz
• AC Current: 29 µA to 20.5 A, 10 Hz to 30 kHz
• Simultaneous Voltage and Current: Up to 20.9 kW equivalent power with phase control to 0.01°
• Dual DC Voltage Channels: Independent sourcing capability
Resistance, Capacitance, and Waveforms:
• Resistance: 0 Ω to 1100 MΩ
• Capacitance: 220 pF to 110 mF
• Waveforms: Sine, square, triangle, truncated sine with harmonic sourcing
Temperature Simulation:
• Thermocouple Types: B, C, E, J, K, L, N, R, S, T, U
• RTD Sensors: Pt 385 (100 Ω, 200 Ω, 500 Ω, 1000 Ω), Pt 3926-100 Ω, Pt 3916-100 Ω, PtNi 385-120 Ω, Cu 427-10 Ω
Oscilloscope Calibration (Optional):
• Bandwidth: Up to 600 MHz
• Edge Rise Time: < 300 ps
• Functions: Leveled sine wave, fast edge, time marker, wave generator, pulse generation, trigger testing, input impedance measurement
Current Extension:
• With 52120A Transconductance Amplifier: 120 A output
• With Accessory Current Coils: Up to 6000 A
Accuracy:
• DC Voltage (1 Year): ±0.005% of setting
• DC Current (1 Year): ±0.01% of setting
• Frequency Uncertainty: < 25 ppm
Interfaces and Power:
• Communications: RS-232, IEEE-488 (GPIB)
• Line Power: Selectable 100 V, 120 V, 220 V, 240 V
• Line Frequency: 47 Hz to 63 Hz
• Line Voltage Tolerance: ±10%
– Key Features
• Calibrates 3.5- and 4.5-digit multimeters with traceable DC accuracy
• Sources multiple waveforms and harmonics for complex signal simulation
• Phase control to 0.01° for power factor and phase angle measurement validation
• Phantom load capability to 20.9 kW for three-phase power equipment testing
• Supports nine thermocouple types plus eight RTD sensor configurations
– Typical Applications
• Calibration of handheld and bench multimeters
• Current clamp verification and calibration
• Panel meter, chart recorder, and data logger adjustment
• Oscilloscope performance validation (with optional module)
• Thermocouple and RTD temperature transducer calibration
– Compatibility & Integration
Remote operation via RS-232 or IEEE-488 GPIB enables integration into automated calibration systems. MET/CAL® Plus software support provides standardized procedures, documentation, and traceability for regulated environments.

















Reviews