The Fluke 732C DC Voltage Reference Standard is a portable, battery-powered voltage source for disseminating and maintaining DC reference standards in primary and secondary calibration laboratories. It delivers three selectable output voltages—10 V, 1 V, and 0.1 V DC—each traceable to national standards and the Fluke Calibration 10 V Josephson Array. Engineered for field portability and laboratory stability, the 732C combines long battery runtime with temperature-stable outputs and comprehensive power options for continuous or mobile operation.
– Technical Specifications
Output Voltages & Stability
• 10 V DC: ±2 µV/V per year (Base models); ±1 µV/V per year (Select models)
• 1 V DC: ±3 µV/V per year
• 0.1 V DC: ±9.8 µV/V per year
• Stability specifications apply at Tcal ±1 °C with IN CAL indicator active
• Outputs referenced to common VCOM binding post
Temperature Performance
• 10 V output temperature coefficient: 0.04 µV/V-°C
• Accounts for drift when operating outside calibration temperature
Power & Battery
• Internal battery: ≥72 hours minimum runtime at 23±5 °C with 0–0.1 mA total output current drain
• External DC input (rear panel): 12 V to 15 V DC, ≥300 mA capacity for indefinite operation
• AC line input: 100 V (90–110 V), 120 V (108–132 V), 220 V (198–242 V), 240 V (216–264 V) at 50/60 Hz
• AC line current at 120 V: typically 0.13 A
• Battery charging time: 10,000 MΩ to earth/chassis ground or AC line, shunted by <1000 pF
• All outputs short-circuit protected indefinitely
• 10 V output: rated for maximum 50 mA continuous at ≤220 V DC
– Key Features
• Three independently accessible output voltage levels
• Selectable Select models offer 2× greater stability at 10 V output
• Long-term stability dominates uncertainty over 30+ day periods
• Field-deployable with minimal power dependencies
• High source impedance isolation for sensitive measurement environments
– Typical Applications
• Primary and secondary reference standard maintenance
• Calibration laboratory voltage dissemination
• Portable on-site calibration verification
• DC voltage standard traceability chain validation
– Compatibility & Integration
The 732C accepts multiple AC line voltages worldwide and supports external 12–15 V DC supplies for laboratory or mobile benchtop integration. Binding post outputs enable direct connection to high-impedance calibration instruments and precision voltmeters.