The HP/Agilent 3562A is a dual-channel dynamic signal analyzer delivering time- and frequency-domain measurements from DC to 100 kHz. Built on FFT technology, it combines network analysis, spectrum analysis, waveform capture, transient analysis, and modulation demodulation in a single platform. The instrument achieves 80 dB dynamic range (or over 130 dB with automatic input ranging), ±0.15 dB amplitude accuracy, and ±0.5° phase accuracy. An integrated signal source generates random noise and sine signals up to ±10 Vpeak, making it ideal for characterizing electronics, mechanical systems, and electromechanical control loops without external stimulus equipment.
– Technical Specifications
Frequency Domain
• Frequency range: DC to 100 kHz
• Resolution: 25.6 µHz (zoom analysis); 801 lines from 2048-point records (linear FFT); logarithmic resolution available
• Measurement range: 150 dB
• Dynamic range: 80 dB (full-scale); >130 dB with automatic input ranging
• Amplitude accuracy: ±0.15 dB ±0.015% (−27 to −40 dBV); ±0.25 dB ±0.025% (−41 to −51 dBV)
• Phase accuracy: ±0.5°; channel match ±0.1 dB, ±0.5° at full scale
Signal Source
• Output: Random noise and sine signals, −10 to +10 Vpeak (AC + DC)
• Output impedance: 50 ± 5 Ω
• Load requirement: ≥10 kΩ, 130 dB achievable with auto-ranging)
• Built-in stimulus source eliminates need for external generators
• Flexible triggering with pre- and post-trigger buffering
• Modulation demodulation for AM/FM/PM signals
– Typical Applications
• Electronics characterization and validation
• Mechanical vibration and modal testing
• Electromechanical control system analysis
• Network analysis via frequency response measurement
• Transient capture and waveform analysis
– Compatibility & Integration
HP-IB interface for external disk drive connectivity and remote control integration.