The HP/Agilent 66319B is a dual-output DC source purpose-built for mobile communications device testing with integrated battery emulation. It characterizes battery operating life and detects product failures early by simulating real-world battery conditions in cellular phones, PDAs, Bluetooth devices, and wireless LAN equipment. Dual outputs enable simultaneous testing of main battery and charger circuits, optimizing bench space and eliminating the need for multiple supplies.
## Technical Specifications
**Electrical Output (per output)**
• Voltage: 0 to 15 V
• Current: 0 to 3 A (continuous); 5 A peak (up to 7 ms)
• Maximum power: 45 W per output
• Voltage ripple and noise (20 Hz to 20 MHz): 1 mV rms / 6 mVpp
• Programming accuracy at 25°C ±5°C: ±10 mV voltage, ±1.33 mA current
• DC readback accuracy: 0.03% + 5 mV (voltage); 0.2% + 0.5 mA at -3 A to +5 A (current)
• Output response time (10/90%): <200 µs
• Voltage settling: <6 ms (2 ms in fast mode)
• Transient voltage dip: 40 mV with <20 µs response time
**Battery Emulation**
• Simulates internal battery resistance effects and wiring voltage drop compensation
• High capacitance compensation: accommodates 50% load current changes
• Remote sensing capability: up to 3 V drop on main output, 4 V on auxiliary output
**Measurement System**
• Dynamic measurement buffer: 4096 points
• Sampling interval: 15 µs variable
• Data acquisition and query processing: 50 ms average
## Key Features
• Dual-output architecture for simultaneous main battery and charger testing
• Programmable output resistance for battery emulation
• Fast transient response (<20 µs) for dynamic load conditions
• Low voltage ripple (1 mV rms) for sensitive device characterization
• Advanced measurement system for current drain analysis across device modes
## Typical Applications
Battery life testing, charger circuit validation, power management IC characterization, mobile device compliance testing.
## Compatibility & Integration
Indoor use rated; Installation Category II, Pollution Degree 2. EMC compliance: 89/336/EEC (ISM 1B). Operating altitude up to 2000 m, maximum humidity 95%.