The HP/Agilent 8722ES is a 40 GHz S-parameter vector network analyzer delivering precise RF and microwave component characterization from 50 MHz to 40 GHz. Integrated synthesized source, test set, and tuned receiver architecture enables accurate magnitude and phase measurements across forward and reverse transmission paths. The instrument maintains 100 dB dynamic range and supports two-port, adapter-removal, and optional TRL calibration for measurement integrity in demanding component validation.
– Technical Specifications
Frequency Range: 50 MHz to 40 GHz
Dynamic Range: 100 dB
IF Bandwidth: 6 kHz to 10 Hz (selectable)
Source Match:
• 0.05–2 GHz: 40 dB
• 2–8 GHz: 40 dB
• 8–20 GHz: 30 dB
• 20–40 GHz: 23 dB
• (Option 400): 40 dB (0.05–0.5 GHz), 40 dB (0.5–2 GHz), 50 dB (2–8 GHz), 50 dB (8–20 GHz)
Directivity:
• 0.05–2 GHz: 42 dB
• 2–8 GHz: 42 dB
• 8–20 GHz: 34 dB, 42 dB
• 20–40 GHz: 26 dB, 38 dB
• (Option 400): 48 dB (0.05–0.5 GHz), 48 dB (0.5–2 GHz), 50 dB (2–8 GHz), 50 dB (8–20 GHz)
Load Match:
• 0.05–2 GHz: 42 dB
• 2–8 GHz: 42 dB
• 8–20 GHz: 34 dB, 42 dB
• 20–40 GHz: 26 dB, 38 dB
• (Option 400): 48 dB (0.05–0.5 GHz), 48 dB (0.5–2 GHz), 50 dB (2–8 GHz), 50 dB (8–20 GHz)
Reflection Tracking:
• 0.05–2 GHz: ±(0.006 dB + 0.03 dB/°C)
• 2–8 GHz: ±(0.029 dB + 0.03 dB/°C), ±(0.010 dB + 0.03 dB/°C)
– Key Features
S-Parameter Measurements: S11, S21, S12, S22; forward/reverse reflection and transmission; receiver signal levels (A, B, R); ratioed measurements (A/R, B/R, A/B)
Parameter Conversion: Z and Y reflection/transmission equivalents; complex S-parameter inverse (1/S)
Display Formats: Cartesian (log/linear magnitude, phase, group delay, SWR, real/imaginary); Smith chart (log/linear magnitude, phase, R+jX, G+jB); polar (magnitude, phase, real/imaginary)
Advanced Analysis: Group delay; time-domain measurements (Option 010); mixer measurements with frequency offset (Option 089); comprehensive marker set with delta, search, bandwidth, and statistics functions
– Typical Applications
Component characterization; filter design validation; amplifier and mixer performance assessment; interconnect and packaging loss measurement; cascade system analysis
– Compatibility & Integration
Option 010 enables time-domain capability. Option 089 adds mixer measurement functionality. Option 400 provides enhanced source match, directivity, and load match across lower frequency bands.