The IOtech TempScan 1000 is a rack-mountable, high-speed scanning instrument for simultaneous acquisition of temperature and voltage data across numerous channels. It measures up to 992 channels when configured with expansion chassis and appropriate scanning modules, supporting standard thermocouple types (J, K, T, E, N, R, S, B, N14, N28) plus custom types, and programmable voltage ranges of ±10V, ±5V, ±1V, and ±100 mV. The system delivers 0.1 °C temperature resolution and voltage resolution down to 3.125 µV in the ±100 mV range. Onboard memory extends from 128 KB to 4 MB for extended data logging, with dual programmable scan rates supporting pre- and post-trigger sampling and event-triggered acceleration. A built-in real-time clock provides time and date stamping for trend monitoring. Communication options include IEEE 488, RS-232/422, and optional RS-232 to USB connectivity via adapter kit. The instrument features 32 TTL digital alarm outputs and 8 TTL-compatible digital inputs for system integration. ChartView software—a Windows-based utility—enables configuration and operation without programming. Channel-to-channel isolation reaches 500 VDC peak with compatible scanning modules. The TempScan 1000 operates on selectable AC power (105-125V or 210-250V) and supports stand-alone operation after disconnection from the host PC.
– Technical Specifications
• Channel Capacity: Up to 992 channels (temperature or voltage) with expansion chassis
• Thermocouple Types: J, K, T, E, N, R, S, B, N14, N28, and custom types with user-defined linearization
• Voltage Ranges: ±10V, ±5V, ±1V, ±100 mV (programmable)
• Temperature Resolution: 0.1 °C
• Voltage Resolution: 3.125 µV (±100 mV range); 312.121 µV (±10V range)
• Memory: 128 KB onboard, expandable to 4 MB
• Channel-to-Channel Isolation: 500 VDC peak
• Real-Time Clock: Built-in; time and date stamping
• Digital I/O: 32 TTL alarm outputs; 8 TTL-compatible inputs
• Power: 105-125V AC or 210-250V AC (selectable)
– Key Features
• Dual programmable scan rates with pre- and post-trigger sampling
• Event-triggered accelerated sampling capability
• DIP switch configuration for RS-232, RS-422, or IEEE 488 protocol selection
• Stand-alone operation independent of host PC
• Flexible connectivity: IEEE 488, RS-232/422 standard; USB via optional adapter
– Interfaces and Communication
• IEEE 488 (standard)
• RS-232/422 (standard)
• RS-232 to USB (optional adapter kit: CA-232-USB-KIT)
• DIP switch rear-panel configuration for protocol and handshake settings
– Typical Applications
• Multi-channel thermocouple acquisition in laboratory environments
• Industrial process monitoring and control integration
• Extended data logging and trend analysis systems
























Reviews