The Kepco ATE150-7DM is a unipolar, linear, programmable DC power supply delivering 500 W across a three-quarter-rack footprint. Built for laboratory, industrial, and system-level test applications, this unit combines precision analog control with digital LCD panel metering for real-time output monitoring. It operates in constant voltage (CV) and constant current (CC) modes with automatic crossover, supported by programmable overvoltage protection and adjustable limiting circuits. The supply accepts local potentiometer control, remote 0-10 V analog programming, and optional IEEE-488 digital interfaces. Two uncommitted internal amplifiers enable signal conditioning and scaling for custom control architectures.
## Technical Specifications
• **Output Range:** 0-150 V, 0-7 A
• **Power:** 500 W
• **Line Regulation (CV):** <0.0005% E_o max.
• **Line Regulation (CC):** <0.01% I_o max.
• **Load Regulation (CV):** <0.01% E_o max.
• **Load Regulation (CC):** <0.02% I_o max.
• **Source Effect (CV):** <0.0005% E_o max.
• **Time Drift (8 h, CV):** <0.005% E_o max.
• **Time Drift (8 h, CC):** 0.01% I_o max.
• **Temperature Coefficient (CV):** <0.01% E_o max.
• **Temperature Coefficient (CC):** 0.02% I_o max.
• **Output Ripple & Noise (CV RMS):** <0.1 mV
• **Output Ripple & Noise (CV P-P):** <1 mV
• **Output Ripple & Noise (CC RMS):** <0.03% I_o max.
• **Output Ripple & Noise (CC P-P):** <0.3% I_o max.
• **Operating Temperature:** 0–55°C
• **Output Impedance (Slow):** R_s = 0.42 mΩ, L_s = 4 μH
• **Output Impedance (Fast):** R_p = 300 kΩ, C_p = 0.3 μF
## Key Features
• Analog control with high gain and zeroable offsets for tracking external signals or resistance inputs
• Automatic CV/CC crossover with no external intervention required
• Digital LCD metering for voltage and current observation
• Programmable analog overvoltage crowbar protection (0-10 V command)
• Optically isolated control signal routing for multi-supply interconnection
• Panel-mounted LED mode indicators
## Typical Applications
• Component and circuit characterization requiring stable, low-noise bias
• Automated test equipment (ATE) control and sequencing
• Research and development bench supply with precision tracking
• Industrial process power and voltage regulation
## Compatibility & Integration
Local control via 10-turn potentiometers. Remote analog programming accepts 0-10 V (voltage) and 0-1 V (current) command signals. Optional SN-series IEEE-488 GPIB interfaces enable digital control and remote configuration. User port with PC-12 mating connector supports jumper-based setup options.