The Keysight/Agilent/HP E1340A is a VXI-based arbitrary function generator replacement board delivering signal generation across DC to 15 MHz with 12-bit resolution and 42 MSa/s sampling. This B-size, single-slot module combines arbitrary waveform generation with standard waveforms (sine, square, ramp, triangle, sin(x)/x, noise, haversine), linear frequency sweep, FSK modulation, and waveform hopping for comprehensive signal conditioning in test systems.
– Technical Specifications
Waveform Generation
• Arbitrary waveforms with 16 kSa memory, segmentable into 4, 8, or 16 kSa blocks
• Up to 4 waveforms in active memory
• Burst mode: 1 to 65,534 cycles or continuous
• Maximum sine frequency: 15 MHz
• Square/ramp/triangle maximum: 1 MHz
• Square wave risetime: 25 ns (typical); jitter: 24 ns
Output Performance
• Amplitude: 10.2 Vp-p maximum into 50 Ω (extendable to 20 Vp-p with external amplifier)
• Output impedance: 50 Ω ± 1%
• DC amplitude accuracy: ± 2.7% of full scale
• AC amplitude accuracy: ± 0.43 dB
• Offset range: ± 5.11 V into 50 Ω
• Sine THD: −60 dB (0–1 MHz), −63 dB (0.1–1 MHz)
• Sine flatness: 0.2 dB to 10 MHz; 0.5 dB to 15 MHz
Frequency Control
• Linear sweep: 0.01 Hz to 15 MHz
• Sweep rate: 0.4 Hz/s to 2.6 MHz/s
• FSK rate: up to 3 kHz
• 42.9 MHz timebase (50 ppm) or external clock input
Integration
• VXI register-based device
• External inputs: clock, burst, gate, waveform hop, FSK
• Supported drivers: I-SCPI Win 3.1, I-SCPI Series 700, C-SCPI LynxOS, C-SCPI Series 700, HP Panel Drivers
• Firmware: Command Module revision A.06 or later
– Key Features
• 12-bit DAC with fine (11-bit) and coarse (0 dB/20 dB) attenuation control
• Configurable waveform looping and modulation via FSK
• Low jitter (24 ns) for timing-sensitive applications
• Multiple input gates for external signal control
– Typical Applications
Component and subsystem testing in VXI chassis, signal conditioning for receiver validation, frequency sweep characterization, and modulated waveform generation for RF and analog circuit verification.
– Compatibility & Integration
Designed for VXI mainframes. Register-based command architecture supports standard SCPI instrument drivers across Windows, LynxOS, and dedicated HP panel control environments. External clock and modulation inputs enable synchronization with test sequencing and stimulus switching.



















Reviews