The Keysight N1913B EPM Series Single-Channel Power Meter delivers accurate average power measurements across DC to 120 GHz with dynamic range from -70 dBm to +44 dBm. Built for bench, rack, and field deployment, this instrument combines a color LCD display with measurement speeds up to 400 readings/second and absolute accuracy of ±0.02 dB (logarithmic) or ±0.5% (linear). The meter expands to three channels via optional USB power sensors and integrates into automated test systems with external trigger capability.
– Technical Specifications
Measurement Performance
• Measurement type: Average power
• Frequency range: DC to 120 GHz (sensor dependent)
• Dynamic range: -70 dBm to +44 dBm (sensor dependent)
• Absolute accuracy: ±0.02 dB (logarithmic) or ±0.5% (linear)
• Relative accuracy: ±0.04 dB (logarithmic) or ±1.0% (linear)
• Measurement speed: Up to 400 readings/second (Fast mode, E-Series sensors); Normal mode 20 readings/second; x2 mode 40 readings/second
• Zero set stability: 0.0000175% (meter only)
• Averaging filter: Selectable 1 to 1024 with auto-averaging noise compensation
• Duty cycle entry: 0.001% to 99.999% for peak power representation
• Limit range: -150.000 dBm to +230.000 dBm (high and low)
Channel Configuration
• Native single-channel operation
• Expandable to three channels via two additional X-Series USB power sensors (U2000 Series, U8480 Series, or U2050/60 Series, excluding peak power measurement)
– Key Features
• Industry-first color LCD display for enhanced data visualization
• Automated frequency and power sweep measurements with optional external trigger in/out
• Internal or external source calibration with reference cal factor adjustment from 1% to 150% in 0.1% increments
• Automatic zeroing when power reference calibrator is switched off
• Code compatibility with legacy Keysight EPM Series meters (E4418B, E4419B) and legacy models 436A, 437B, 438A with N191xB-200 options
– Sensor Compatibility
• Compatible with Keysight E9300 Series sensors (excluding E9320), E4410 Series, N8480 Series, and discontinued 8480 models
• Supports X-Series USB power sensors for multi-channel expansion
– Typical Applications
Automated test systems, bench and field power measurements, RF/microwave characterization, duty cycle analysis via peak power calculation.















Reviews
There are no reviews yet.