The NI SCXI-1162HV is a 32-channel optically isolated digital input module for SCXI chassis-based data acquisition systems. It safely measures digital signals in high common-mode voltage and ground-loop environments by providing individual optical isolation on each channel. The module protects DAQ systems and connected equipment from voltage spikes and surges while accepting digital logic levels up to 240 VAC/VDC across diverse industrial and laboratory applications.
– Technical Specifications
Input Characteristics
• Input signal range: Digital logic levels up to 240 VAC/VDC
• DC input threshold: ≥2 V or ≤−2 V registers as logic high
• AC input thresholds: 50/60 Hz at ≥10 VAC (logic high); 1 kHz at ≥2 VAC (logic high); 1 kHz at ≤±1 V peak (logic low)
• Input current limiting: 1 mA
• Transient overvoltage protection: 2,300 Vrms
• Isolation voltage: 300 Vrms per bank, Cat. II
Timing and Data Rate
• Propagation delay (parallel mode): 1.5 ms
• Serial data rate: 1,000 words/s (system dependent; measured on 500 MHz Pentium III)
Physical
• Dimensions: 1.2 × 6.8 × 8.0 inches
• Weight: 0.96 lbs
• Front connector: 48-pin DIN C male
• Rear connector: 50-pin male ribbon cable
– Key Features
• 32 independently isolated input channels
• Dual operating modes: serial (multiplexer mode) and parallel mode
• Operation in industrial environments
• Integrated transient protection rated to 2,300 Vrms
– Compatibility & Integration
DAQ Devices
• Compatible with NI SCXI chassis systems
• Programmable via SCXIbus by any compatible DAQ device
• Direct parallel cabling to DIO boards: 6503 (DIO-24), 6533 (DIO-32HS), 6534, 6508 (DIO-96)
• Works with MIO Series boards, SCXI-1200 module, Lab series boards, DIO-24, DIO-32F, and DIO-96
Software
• NI-DAQ 7 driver software
• LabVIEW, LabWindows, and NI-DAQ programming support
Note: Product is discontinued by manufacturer.