The Sencore LC102 AUTO-Z is a microprocessor-controlled capacitor and inductor analyzer engineered for rapid identification of defective components in service, maintenance, and incoming inspection environments. This portable instrument automatically measures capacitor value, leakage current, equivalent series resistance (ESR), and dielectric absorption, while inductors are analyzed for value and quality using patented test methods. Results display on a clear LCD in GOOD/BAD format against user-defined or EIA standards, with automatic lead zeroing that compensates for test lead capacitance, resistance, and inductance to ensure accuracy on small components.
– Technical Specifications
Capacitor Value
• Range: 1.0 pF to 19.99 F, fully autoranged across 12 ranges
• Accuracy: ±1% ±1 pF ±1 digit up to 1990 µF; ±5% ±0.1% of full scale from 2000 µF to 19.99 F
• Resolution: 0.1 pF (lowest) to 0.01 F (highest)
• Test method: Dynamic RC time constant measurement with voltage-dependent charging (1.5 MΩ path for 0–0.002 µF; 15 kΩ path for 0.002–2 µF; 60 mA constant current for 2–2000 µF; 416 mA for 2000 µF–19.99 F)
• Maximum test voltage: 1.75 V for capacitors > 2 µF
Equivalent Series Resistance (ESR)
• Range: 0.10 Ω to 1999 Ω, fully autoranged
• Accuracy: ±5% ±1 digit
• Resolution: 0.01 Ω to 1 Ω across three ranges
Leakage Current
• Range: 0.01 µA to 19.99 mA, fully autoranged
• Accuracy: ±5% ±1 digit
• Resolution: 0.01 µA to 0.01 mA across four ranges
Dielectric Absorption
• Range: 1% to 100%
• Accuracy: ±5% of reading ±1 digit
• Measurement range: 0.01 µF to 19.99 F
• GOOD/BAD threshold: ±15% variation for electrolytics
Leakage Test Voltage
• Range: 1.0 V to 999.9 V in 0.1 V steps
• Accuracy: +0% to -5%
• Short circuit current limit: < 900 mA
• Continuous power: 6 W ±10%
– Key Features
• Patented dielectric absorption and double-layer lytics testing
• Automatic lead zeroing eliminates test lead errors on precision measurements
• Voltage detection with automatic test button disable prevents damage from external voltages
• Fused input protection
• Fully autoranged measurements eliminate manual range selection
– Typical Applications
• Component-level service and repair
• Manufacturing incoming inspection
• Laboratory batch testing
• Field maintenance diagnostics




















Reviews