The HP 70950A Optical Spectrum Analyzer (OSA) module integrates with the HP 70004A mainframe to deliver precision spectral analysis across the 600 nm to 1700 nm range. This diffraction-grating-based instrument measures optical power versus wavelength with high amplitude and wavelength accuracy while remaining polarization insensitive. The double-pass monochromator design achieves high dynamic range performance comparable to double-monochromator systems, combined with single-monochromator sensitivity. Ideal for characterizing optical components, amplifiers, LEDs, DFB lasers, and Fabry-Perot lasers, the HP 70950A operates as the HP 71450A when paired with its mainframe in the Hewlett-Packard Modular Measurement System (MMS) platform.
– Technical Specifications
• Wavelength Range: 600 nm to 1700 nm
• Resolution Accuracy: Greater than 0.5 nm (1250 nm to 1600 nm range), ±20% accuracy
• Maximum Input Power: +20 dBm per 5 nm bandwidth; +30 dBm total
• Operating Temperature: 0°C to 55°C
• Humidity: <95% relative humidity
• Polarization: Insensitive
– Key Features
• Double-pass monochromator design for high dynamic range with single-monochromator sensitivity
• High amplitude and wavelength accuracy
• Softkey menu control via mainframe display
• Auto-measure and auto-align functions
• Rigorous environmental qualification for shock and vibration
– Typical Applications
• Optical component and amplifier characterization
• LED and DFB laser spectral analysis
• Fabry-Perot laser measurement
• EDFA testing (with Option 051 or 052)
– Compatibility & Integration
Designed for the HP Modular Measurement System. Operates with HP 70004A mainframe. Optional configurations include programmable current source (Option 001), white light source (Option 002), 9 µm input connector (Option 009), and EDFA test personalities. Input connector options available separately: FC/PC (81000FI), FC/APC (81000NI), SC/PC (81000KI), and ST (81000VI).