The Tektronix DM5110 is a 4.5-digit programmable digital multimeter designed for precise electronic measurements in automated test systems and laboratory environments. This modular plug-in for the TM5000 system delivers IEEE-488.1 (GPIB) programmability, enabling remote control and seamless integration into larger test configurations. With autoranging capability, multiple measurement modes, and built-in comparison functions, the DM5110 handles demanding measurement tasks from DC voltage to temperature across a wide dynamic range.
– Technical Specifications
Measurement Functions & Ranges:
• DC Voltage: up to 1000 V, 10 mV resolution
• AC Voltage (True RMS): up to 500 V, 10 mV resolution, 50 kHz frequency response
– 20–100 Hz: ±(0.8% reading + 0.05% F.S.)
– 100 Hz–10 kHz: ±(0.3% reading + 0.05% F.S.)
– 10–20 kHz: ±(0.6% reading + 0.05% F.S.)
– 20–50 kHz: ±(1.0% reading + 0.05% F.S.)
• DC Current: up to 2000 mA, 10 nA resolution
• AC Current (True RMS): up to 2000 mA, 10 nA resolution
• Resistance: up to 20 MΩ, 10 mΩ resolution
• Temperature: −62 °C to +240 °C
• dBm/dBV calculations via software
Display & Indicators:
4.5-digit LED display (±19,999 counts) with NULL, HOLD, AUTO, ADDR, and REM indicators.
Basic DC Accuracy: ±0.05% (DC Voltage 2 V range: ±(0.1% reading + 0.02% F.S.))
– Key Features
• Autoranging with manual range override
• IEEE-488.1 GPIB interface (SH1, AH1, T6, L4, SR1, RL1, PPO, DC1, DT1, E1 functions)
• Null offset measurement capability
• Hold mode for value retention
• Triggered measurement operations
• Compare mode with user-selectable HI/LO limits
• Beeper alarm for limit exceedance
• FAST (3.5-digit) and NORMAL (4.5-digit) measurement modes
• 50/60 Hz line frequency measurement modes
• Front/rear panel input selection for LOW and Volts/Ω/Temp inputs
– Typical Applications
Automated testing, data acquisition, production environments, and laboratory measurement systems requiring programmable control and high measurement accuracy.
– Compatibility & Integration
Modular TM5000 system plug-in with full IEEE-488.1 GPIB programmability for system integration.