The Teledyne LeCroy CP030A is a high-sensitivity current probe engineered for precise AC, DC, and impulse current measurements up to 30 A RMS without circuit interruption or shunt insertion. Operating across a 50 MHz bandwidth with 7 ns risetime, the probe delivers 1 mA/div sensitivity and ±1% accuracy (±10 mA) at currents to 30 A, making it ideal for power supply characterization, motor control analysis, and automotive electronics diagnostics.
– Technical Specifications
• Maximum Continuous Input Current: 30 A RMS
• Maximum Peak Current: 50 A (pulses <10 µs)
• Bandwidth: 50 MHz
• Sensitivity: 1 mA/div
• Amplitude Accuracy (≤30 A): ±1% of reading ±10 mA
• Amplitude Accuracy (≤50 A): ±2% of reading ±10 mA
• Risetime: 7 ns typical
• AC Noise: <2.5 mA
• Maximum Circuit Voltage: 300 V CAT I (insulated wire)
• Maximum Conductor Size: 5 mm
• Cable Length: 2 m
• Coupling Modes: AC, DC, GND
• DC Input Current (0.1 V/A sensitivity): 30 A
• DC Input Current (1 V/A sensitivity): 5 A
– Key Features
• Hall effect and transformer hybrid technology enables simultaneous AC/DC measurement without circuit breaks
• Small jaw design accommodates dense PCB layouts and tight conductor access
• ProBus active probe interface integrates power and communication with compatible oscilloscopes
• Automatic waveform scaling displays current in Amperes and power in Watts directly on scope display
• Degauss and Autozero functions controllable from oscilloscope interface
– Compatibility & Integration
The CP030A interfaces via ProBus and is compatible with Teledyne LeCroy MAUI oscilloscopes running Windows XP Professional or later with firmware version 7.7.1.Xa or greater. Not compatible with WaveSurfer 3000 series instruments.