The Teledyne LeCroy HDA125-09-LBUS High-Speed Digital Analyzer is a modular accessory that extends oscilloscope capabilities for mixed-signal testing and digital design validation. It delivers 12.5 GS/s sampling across up to 18 digital input channels with 80 ps timing accuracy, enabling capture and analysis of high-speed interfaces such as DDR4 up to 6 Gb/s. The analyzer connects via LBUS or ChannelSync interface for precise timebase synchronization with compatible Teledyne LeCroy oscilloscopes, supporting cross-triggering between instruments. Digital waveforms display directly on the oscilloscope, time-correlated with analog traces, while bus decoding consolidates multiple digital signals into decoded displays.
– Technical Specifications
• Digital Sampling Rate: 12.5 GS/s on all 18 input channels
• Timing Accuracy: 80 ps
• Maximum Digital Signal Capture: Up to 6 Gb/s
• Digital Channels: Up to 18 (model -09 indicates 9-channel leadset)
• Input Sensitivity: 150 mV minimum signal swing
• Threshold Accuracy: Three times better than competing mixed-signal instruments
• Hysteresis Control: Adjustable 50 mV to 600 mV range
• Probe Tip Impedance: 110 kΩ, 0.12 pF differential
• Leadset Bandwidth: 3 GHz
• QuickLink Probe Tip Bandwidth: 8 GHz
• Data Interface: USB 3.1 (USB 2.0 supported at reduced speed)
– Key Features
• QuickLink differential probing system with 9-inch solder-in leads
• Ultra-low probe loading for minimal circuit impact
• Adjustable hysteresis thresholds for threshold validation
• Time-correlated digital and analog waveform display
• Bus decoding of parallel digital signals
• Digital triggering on parallel bus patterns
• Compatible with Teledyne LeCroy WaveLink differential analog probes
– Typical Applications
• Debugging and validating high-speed digital designs
• Mixed-signal analysis on DDR4 and similar interfaces
• Parallel bus protocol monitoring and triggering
• Timing correlation between digital and analog domains
– Compatibility & Integration
Compatible Oscilloscopes: WaveMaster 8Zi family, WavePro 7Zi family, WaveRunner 6Zi series, HDO9000, LabMaster 9Zi, LabMaster 10Zi-A (requires LBUS or SYNC connector and software version 8.1.x.x or higher). Not compatible with WaveRunner 8000, HDO, MDA, WaveSurfer 3000, or legacy models.